Molecular Devices·Lectores de microplacas
SpectraMax M3
SpectraMax M3 — lector de microplacas de Molecular Devices (Dimensions (H): 8.6 in.).
8.6 in.Dimensiones
22.8 in.Dimensiones
15.3 in.Dimensiones
Descripción
Conoce el equipo
SpectraMax M3 es un(a) lector de microplacas de Molecular Devices, indicado para ensayos de absorbancia, fluorescencia y luminiscencia, cuantificación de proteínas y ácidos nucleicos, ensayos celulares. Especificaciones destacadas: Dimensions (H): 8.6 in.; Dimensions (W): 22.8 in..
Aplicaciones ideales
Resuelve tu aplicación
Ensayos de absorbancia, fluorescencia y luminiscencia
Cuantificación de proteínas y ácidos nucleicos
Ensayos celulares
Especificaciones técnicas
Especificaciones
del SpectraMax M3

- Dimensions (H)
- 8.6 in.
- Dimensions (W)
- 22.8 in.
- Dimensions (D)
- 15.3 in.
- Dimensions (H)
- 22 cm
- Dimensions (W)
- 58 cm
- Dimensions (D)
- 39 cm
- Weight
- 36 lbs.
- Weight
- 16.4 kg
- Power consumption
- < 420 watts
- Power source
- 100–240 VAC, 3.5 A, 50/60Hz
- Robotic-compatible
- Yes
- Plate formats
- 6, 12, 24, 48, 96, 384 wells
- Light source
- Xenon Flash Lamp (1 joule/flash)
- Detectors
- 2 photomultiplier tubes (PMT)
- Shaker time
- 0 to 999 seconds
- Temp. control
- 2°C above ambient to 60°C
- Temp. uniformity
- < 1°C
- Temp. accuracy
- ±1°C
- Endpoint reading
- All modes
- Kinetic reading
- All modes
- Spectral scanning
- All modes
- Well scanning
- Abs, FI, TRF, Lum
- Absorbance read time (96 wells)
- 0:18 minutes:seconds
- Absorbance read time (384 wells)
- 0:49 minutes:seconds
- Fluorescence Intensity read time (96 wells)
- 0:17 minutes:seconds
- Fluorescence Intensity read time (384 wells)
- 0:48 minutes:seconds
- Luminescence read time (96 wells)
- 2:00 minutes:seconds
- Luminescence read time (384 wells)
- 7:00 minutes:seconds
- Wavelength range (Absorbance)
- 200–1000 nm
- Wavelength selection (Absorbance)
- Monochromator, tunable in 1.0 nm increments
- Wavelength bandwidth (Absorbance)
- ≤ 4.0 nm
- Wavelength accuracy (Absorbance)
- ±2.0 nm
- Wavelength repeatability (Absorbance)
- ±0.2 nm
- Photometric range
- 0–4.0 OD
- Photometric resolution
- 0.001 OD
- Photometric accuracy (microplate)
- < ±0.006 OD ±1.0%, 0–2 OD
- Photometric accuracy (cuvette)
- < ±0.005 OD ±1.0%, 0–2 OD
- Photometric precision
- < ±0.003 OD ±1.0%, 0–2 OD
- Stray light
- < 0.05%
- Wavelength range (Fluorescence Intensity)
- 250–850 nm
- Wavelength selection (Fluorescence Intensity)
- Monochromators, tunable in 1.0 nm increments
- Bandwidth (EX, EM) (Fluorescence Intensity)
- 9 nm, 15 nm
- Optimized sensitivity (Fluorescence Intensity)
- ≤ 1 pM fluorescein in 96 wells, ≤ 1.5pM in 384 wells
- Guaranteed sensitivity (Fluorescence Intensity)
- < 5 pM fluorescein in 96 wells or cuvette, < 20 pM in 384 wells
- Wavelength range (Luminescence)
- 250–850 nm
- Optimized sensitivity (Luminescence)
- ≤ 43pM ATP in 96 wells
- Guaranteed sensitivity (Luminescence)
- ≤ 75pM ATP in 96 wells
- Dynamic range (Luminescence)
- > 6 decades
- Cross-talk (Luminescence)
- < 0.3%
Equipos relacionados
También te puede interesar
¿Te interesa el SpectraMax M3?
Cuéntanos tu aplicación y un especialista de Inolab te ayuda con la configuración, la demostración y la cotización.